6.2.1. IDENTIFICATION
Syntax:
*IDN?
Form:
Query
Description:
Query instrument identification. Response gives instrument model, serial
number and firmware version.
Parameter:
-
RST value:
-
Example:
Answer: GFTy/MIPSI,MOD745-20,SN54001/000000,V1.0
Instrument model: MOD745-20
Serial number: 54001
Firmware version: 1.0
6.2.2. DELAY
Syntax:
DELAY T<n>,<D>
DELAY? T<n>
Form:
Set & Query
Description:
Query delay of channel T<n> or set the delay of channel T<n> to <D>
picosecond (relative to T0 channel)
Parameter:
<n>: Channel number (1 to 20)
<D>: Delay value (in picosecond)
RST value:
Last value set
Range:
0 to 9 999 999 999 999 ps
Example:
Set 1 ns delay on channel 2: DELAY T2,1000
Query delay on channel 2: DELAY? T2 (Answer: :DELAY T2,1000)