Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS Page -20 / 36 -
LEVEL
Syntax: LEVEL START|STOP,<level>
LEVEL? START|STOP
Form: Set & Query
Description: Set or query channel trigger level.
Command
Parameter:
START set level for START channel
STOP set level for STOP channel
<level> level between 0.5 and 5.0 volts
Specified limit 0.5V <= level <= 5.0V
Example: LEVEL START,2.8
LEVEL? START=> :LEVEL START,2.8
MODE
Syntax: MODE SING|REP
MODE?
Form: Set & Query
Description: Set or query mode.
Command
Parameter:
SING set single measurement
REP set repetitive measurement
Query response :MODE SING single measurement
:MODE REP repetitive measurement
Example: MODE SING
MODE? => :MODE SING