in volts as units.
Note
The returned value is 0, and the unit is volts.
[:SENSe]EVENt3:SLOPe?
Queries which edge of channel 3 is used to define an event.
Return
A sequence of ASCII-encoded bytes, POS.
Sets or queries
Gate
for Period, frequency and frequency ratio measurements. For more
information about t
he front panel gating settings and the interface command
parameters, please refer to the FREQ table in Chapter 5.4.
[:SENSe]:FREQuency:ARM[:STARt]:SLOPe POSitive NEGative
Sets or queries the slope of the external start arm signal used in external arming for
Period, frequency and frequency ratio measurements.
Return
A sequence of ASCII-encoded bytes. POS mean positive and NEG means negative.
Notes
This will be set to Positive after the counter has been turned off or a remote interface
reset.
This command only applies when [:SENSe]:FREQ:ARM[:STAR]:SOUR EXT is used.
[:SENSe]:FREQuency:ARM[:STARt]:SOURce IMMediate EXTerna
Sets or queries the start arm for period, frequency and frequency ratio measurements.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means
External.
Note
This will be set to Immediate after the counter has been turned off or a remote
Vedere la pagina 139